产品简介
WIKA,MG-1 氧用压力变送器
WIKA,MG-1 氧用压力变送器
产品价格:¥5000
上架日期:2013-10-09 10:54:19
产地:德国
发货地:北京
供应数量:不限
最少起订:1件
浏览量:259
资料下载:暂无资料下载
其他下载:暂无相关下载
详细说明

    Electronic
    pressure measurement

    Pressure transmitter for medical gases
    Model MG-1



                         Pressure transmitter model MG-1

     

        Applications
        Distribution and storage of medical gases
        Oxygen treatment for patients in hospitals, at home and in ambulances

       Special features
        Measuring ranges from 0 … 6 to 0 … 16 bar and from 0 … 200 to 0 … 400 bar
        Output signals 4 … 20 mA, DC 0 … 10 V, DC 0 … 5 V,
           DC 1 ... 5 V, DC 0.5 … 4.5 V ratiometric
        Oxygen clean in accordance with international standards
        Available in four levels of cleanliness
        Three packaging variants

     

        Debion

        The model MG-1 pressure transmitter has been developed for the measurement of pressure with medical gases and for oxygen applications. Only materials that are suitable for oxygen applications are used.

        In order to ensure the required level of cleanliness,any contamination of the components during production is avoided. The model MG-1 is manufactured under controlled conditions, then it is marked for use in oxygen applications and packed with special care.

        In line with international directives, different levels of cleanliness,different packaging and different instrument markings are available.

        The MG-1 pressure transmitter offers a solution which is optimised for your application.

     

        Measuring ranges



        The given measuring ranges are also available in kg/cm2, MPa and kPa.
        Vacuum and +/- measuring ranges are also available.
        Vacuum resistance
        Yes

     

        Output signals



        Accuracy


       Accuracy at reference conditions
          ≤ ±2 % of span

        Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to measured error per IEC 61298-2).
        Calibrated in vertical mounting b with process connection facing downwards.

       Non-linearity
          ≤ 0.5 (BFSL per IEC 61298-2)

       Temperature error
          ≤ ±2.0 % of span

       Long-term drift (per IEC 61298-2)
          ≤ 0.3 % of span/year

        Settling time
          ≤ 2 ms

       Voltage supply
        The permissible power supply depends on the corresponding output signal.



        Total current consumption maximum 10 mA (except for 2-wire signals)

     

       Operating conditions
       

        Ingress protection (per IEC 60529)
          IP 67

        The stated ingress protection only applies when plugged in using a mating connector that has the appropriate ingress protection.

       Vibration resistance
        20 g (20 ... 2,000 Hz, 2 h) per IEC 60068-2-6 (vibration under resonance)

       Shock resistance
        40 g (6 ms) per IEC 60068-2-27 (mechanical shock)

       Temperatures

       Reference conditions (per IEC 61298-1)

       Temperature
          15 ... 25 °C

        Atmospheric pressure
          860 … 1,060 mbar (665 ... 800 mmHg)

        Humidity
          45 ... 75 % relative, non-condensing

        Power supply
          DC 24 V
          DC 5 V with ratiometric output

        Mounting b
          as required

     

        Process connections


       Electrical connections
          Short-circuit resistance
          S+ vs. 0V

       Reverse polarity protection
          UB vs. 0V

        Insulation voltage
          DC 500 V

        Connection diagrams




       Cleanliness specifications


        Packaging
        For the individual levels of cleanliness of the model MG-1 the following types of packaging are available.



        Materials

        Wetted parts
          Process connection from stainless steel 316L and 13-8 PH
          Sealing ring from FKM (if available)

       Non-wetted parts
          Case from stainless steel 316L
          Electrical connection from highly resistant, glass-fibre reinforced plastic PBT GF 30


        Approvals, directives and certificates
     

        CE conbity
          EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and immunity (industrial application)
          Pressure equipment directive 97/23/EC

     

        Dimensions in mm
        Pressure transmitter


        Process connections





    Ordering inbation
    Model / Measuring range / Output signal / Electrical connection / Process connection / Level of cleanliness / Type of packaging

     

    欢迎致电咨询定购WIKA 氧用压力变送器 MG-1WITH M12*1 connector 压力变送器

    电话:010-51669912 56291930 15910733293

    传真:010-51669912

    QQ:823377546 1491421115 2636968574

    邮箱:sales@zoriver.cn

    公司官网:http://www.zoriver.cn                http://www.zoriver.com.cn